78
Operation
Display
Description
locked
The administrator lock is closed.
open
The administrator lock is open.
6.4.6.2 Cod.A – enter / change administrator code
By entering the correct code (hexadecimal characters) here it is
possible to open the administrator lock. After opening the
administrator lock, the existing code can be changed.
Individual characters of the code can have the values 0 to 9
and A to F.
6.4.6.3 St.PL – Calibration lock: State/close
The calibration lock must be opened in order to modify certain
measurement parameters. The calibration lock can only be opened
by pressing the calibration switch on the CPU board. An open
calibration lock can be closed by pressing the "Enter" key and
selecting the "locked" option. The calibration lock should be closed
when in normal operating mode.
Display
Description
locked
The calibration lock is closed.
open
The calibration lock is open.
In this case, the "P" will also flash in the upper left-
hand side of the display (see chapter 6.3.4
"Meaning of status symbols")
6.4.6.4 Menu - Selection of the display menu
In an as-delivered condition, the display of the DL230 has the
following five registers: "Main", "Cust.", "Admin.", "Serv." and "Ctrl.".
Registers can be displayed and hidden for certain purposes using
the "Menu" value.
Summary of Contents for DL230
Page 1: ...Operating manual Data Logger DL230 ...
Page 8: ......
Page 105: ...Appendix 105 English manual 73 022 387 ...
Page 106: ...106 Appendix 9 2 EC Declaration of Conformity ...
Page 107: ...Appendix 107 9 3 ATEX Type Examination Certificate 9 3 1 Associated apparatus for Zone 0 1 ...
Page 108: ...108 Appendix ...
Page 109: ...Appendix 109 ...
Page 110: ...110 Appendix ...