Maintenance
91
When using one size D battery, the value 13.0 Ah
should be entered for "Bat.C". When using 2 batteries,
the value 26.0 Ah must be entered.
Once you have changed all of the digits, press the ENTER key in
order to confirm your entry. The input value can be deleted by
pressing the ESC button.
To close the administrator lock move the cursor to the "Admin"
register and to the "St.AS" value (Status of administration lock)
via the following path:
Admin.
Device settings
Access
St.AL
Press the ENTER key
"open" flashes.
Move the cursor with the arrow keys
or
in order to change
the value to "locked".
After making the change, press the ENTER button to confirm the
input. The input value can be deleted by pressing the ESC
button.
7.2.3 Display remaining battery power
The remaining battery power is calculated separately
from the consumed power (which is measured) and
from the anticipated future consumption (giving the
theoretical remaining battery power). Therefore, for
applications which are very power-consuming, the
remaining battery power may drop quicker than is
shown on the battery power display.
The remaining battery power will automatically be
recalculated after entering a new battery capacity
Bat.K (see above). The value cannot be edited.
Move the cursor to the "Serv." register and to the "Bat.R" value
(remaining battery power) via the following path:
Admin
Batteries
Bat.R
After entering the new battery capacity, a value of
at least 96 months will be displayed in standard
measurement conditions (see chapter 3.2.1) for
"Bat.R".
Summary of Contents for DL230
Page 1: ...Operating manual Data Logger DL230 ...
Page 8: ......
Page 105: ...Appendix 105 English manual 73 022 387 ...
Page 106: ...106 Appendix 9 2 EC Declaration of Conformity ...
Page 107: ...Appendix 107 9 3 ATEX Type Examination Certificate 9 3 1 Associated apparatus for Zone 0 1 ...
Page 108: ...108 Appendix ...
Page 109: ...Appendix 109 ...
Page 110: ...110 Appendix ...