3.4.3 Humidity channel setup
The default EU (Engineering Unit) range for all voltage/current inputs (as
described in 4.4.2), is 0/1V = 0/100% which is suitable for probes such as
the EplusE EE08 and Rotronic Hygroclip S3/HC2. These probes plug into
any voltage/current input selected using the CHANNEL SETUP function.
Press and hold
set
for 3 seconds to change an unused voltage/current
channel to the default EU range.
N.B.
It is essential that the ‘sensor on’
time is set to 1 second or greater. Please see 4.10.4 for further
details.
3.5 Start logging
Press
function
until the following is displayed: -
L O G o f f
u s e d 0
Now press and hold
set
to start logging: -
L O G o f f
H O L D T O S T A R T
The display shows that logging has started and will turn off within 10
seconds to conserve power. (Some memory slots will be used up
immediately in order to store the header file.)
L O G o n
u s e d 1 4