14
The number of separate runs already recorded (out of a maximum of 99)
is displayed in the last screen. If the logger is logging, then the current run
number is displayed:
L O G o f f
r u n 3
4.1.3 Start/stop logging
If you have just unpacked your datalogger and have not previously used
this model then please go to section 3. Whenever you first enter the LOG
function you can press and hold
set
to start and stop logging: -
L O G o f f
u s e d 1 2 3 4 5
L O G o n
u s e d 1 2 3 4 5
N.B. you can only start and stop logging when the display shows either of
these screens.
4.1.4 How to clear the memory
The procedure for clearing the memory has been designed to reduce the
risk of the user accidentally clearing the memory. It is not possible to clear
the memory while either logging or waiting to log. It is only possible to
clear the memory using the following sequence:
Press function to wake up the datalogger and press select within 10
seconds to obtain either of the following displays: -
L O G o f f
f r e e 1 2 3 4 5
L O G o f f
r u n 3
Now you can clear the memory by pressing and holding
set
for 2 seconds:
L O G o f f
H O L D T O C L E A R