20
Please note that the logging/sampling interval cannot be changed while
logging is in progress or while logging is waiting to commence:
I N T l o g 1 2 : 0 0 : 0 0
L O G G I N G
4.7 The OUTPUT function
This function is used to output data to a computer using the following
programs: -
Data transfer and setup
Darca Lite
Data transfer and analysis
Darca Plus / Squirrelwise
Data transfer, direct/remote setup and analysis
Darca Plus
Once you enter OUTPUT the display will clear and the datalogger will
quickly enter sleep mode, which is required prior to active communication.
The baud rate is automatically selected by the program that is used. Eltek
Squirrels with serial numbers 6000 and above will work at up to 38400
baud. During downloading, the number of blocks to be downloaded is
displayed and this number changes after each successful transmission of
a block. While not actively communicating,
“idle” is displayed. An example
of some of the displays which may be seen during the downloading
process are shown below:
O U T P U T 9 6 0 0 b a u d
c o m m s b l o c k 6
O U T P U T 9 6 0 0 b a u d
i d l e
O U T P U T 9 6 0 0 b a u d
c o m m s b l o c k 5
The number of blocks transmitted will decrement until the display clears to
indicate that all the data have been transferred. Please note that
downloading of data can be achieved while logging is in progress and that
the following minimum logging intervals are permissible, corresponding to
the appropriate baud rates: -