5.3 Wiring up sensors to voltage channels
Individual plug-in terminal blocks with cable restraints are provided to
make the connections to the inputs on the datalogger. The positive output
from the sensor should be connected to the
‘v’ input and the negative to
the
‘-’ input:
c
v
o
c
v
sensor
output
sensor output
sensor
power
sensor output
v
v
c
o
sensor
output
sensor
power
Note: if the cable used has an independent screen, this may be connected
to the ground terminal, which is marked
‘o’.
5.4 Wiring up sensors to current channels
Individual plug-in terminal blocks with cable restraints are provided to
make the connections to the inputs on the datalogger. The connections
are the same as for voltage with the addition of a link between the
‘c’ and
‘v’ inputs. This link is essential and must be provided by the user:
c
v
o
c
v
sensor output
sensor output