63
13. Technical data
13.1 Inputs, ranges and resolutions
Input type
Available ranges
Resolution
Thermistor temperature (U)
-50 to 150°C
-58 to 302°F
See below
Best resolution
-5 to 75°C 0.05°C
-25 to 5°C & 75 to 100°C 0.1°C
-40 to -25°C & 100 to 125°C 0.2°C
Worse resolution
-50 to -40°C & 125 to 150°C 0.4°C
Thermocouple temperature (T)
-200 to 200°C
-328 to 392°F
0.1°C
Impedance: 100k
W
-200 to 350°C
-148 to 662°F
0.2°C
Thermocouple temperature (K)
-200 to 200°C
-328 to 392°F
0.1°C
Impedance: 100k
W
-200 to 450°C
-328 to 842°F
0.2°C
-200 to 1200°C
-328 to 2192°F
0.5°C
PT100 temperature (P3/P4)
-200 to 100°C
-328 to 212°F
0.1°C
-200 to 400°C
-328 to 752°F
0.2°C
-200 to 600°C
-328 to 1112°F
0.25°C
DC voltage
0 to 50mV
-25 to 25mV
50µV
Impedance: 1M
W
0 to 100mV
-50 to 50mV
50µV
0 to 200mV
-100 to 100mV
50µV
0 to 500mV
-250 to 250mV
0.5mV
0 to 1V
-500 to 500mV
0.5mV
0 to 2V
-1 to 1V
0.5mV
0 to 5V
-2.5 to 2.5V
5mV
0 to 10V
-5 to 5V
5mV
0 to 20V
-10 to 10V
5mV
DC current
4 to 20mA (as 0 to 100%)
0.05%
Impedance: 10
W
0 to 20mA
10µA
Pulse count
0 to 65,000
1
Impedance: 1M
W
0 to 650,000
10
Digital
0 to 255
State
Impedance: 1M
W
00000000 to 11111111