18
BA-en-2087-2207_Static Combi Cleaner SCC-P
5.2 Inspection of the protective resistors - contact protection
The function and the appearance of the protective resistors must be
inspected at regular intervals. The inspection intervals are specified in
the accident prevention regulations, as amended (e.g. in Germany DGUV
V3).
The function of the series resistors must be checked using a suitable
measuring device. The test voltage must be 1,000V. The measured resis-
tance between the high-voltage connection and the individual emission tip
must not fall below 80 MOhm and not exceed 120 MOhm.
Please additionally observe the operating instructions for the ion blower
nozzles R36, BA-en-2043.
Summary of Contents for SCC-P
Page 1: ...Operating Instructions Static Combi Cleaner SCC P BA en 2087 2207 Z 118074y_1 ...
Page 2: ...2 BA en 2087 2207_Static Combi Cleaner SCC P ...
Page 4: ...4 BA en 2087 2207_Static Combi Cleaner SCC P ...
Page 21: ...BA en 2087 2207_Static Combi Cleaner SCC_P 21 8 Dimensions Fig 6 Dimensions Z 11805y_3 ...
Page 22: ...22 BA en 2087 2207_Static Combi Cleaner SCC P ...
Page 23: ......