6. Operation

To operate the device, the transparent cover must be removed by undoing 

the two screws.

6.1. Switching on and off

Briefly pressing the button once switches the device on. Pressing the button 

again (also briefly) switches the device off again. .

 

 

6.2. Adjusting the operating temperature

To display the conductivity measurement correctly, the maximum anticipated 

operating temperature of the system must be set.

PIN setting

Max. operating temperature

PIN 1

PIN 2

PIN 3

25 °C

0 (OFF)

0 (OFF)

x (any)

20 °C

0 (OFF)

1 (ON)

x (any)

40 °C

1 (ON)

0 (OFF)

x (any)

60 °C

1 (ON)

1 (ON)

x (any)

6.3. Intervals between measuring cycles

The intervals between the measurements can be adjusted using PIN 3.

 

Position «ON» means that the conductivity is measured every 20 seconds. Po

-

sition «OFF» means that a measurement is carried out every 2 seconds. The 

PIN 1-3

Button

Summary of Contents for PUROTAP i-control-21

Page 1: ...www elysator com Installation Function Operation Maintenance Operating instructions PUROTAP i control 21s...

Page 2: ......

Page 3: ...Specification 4 1 Device description 6 4 2 Dimensions 7 4 3 Performance data 7 5 Installation 8 6 Operation 6 1 Switching on and off 9 6 2 Adjusting the operating temperature 9 6 3 Intervals between...

Page 4: ...4 1 Safety instructions Not drinking water Maximum 60 C Maximum 4 bar pressure Dust proof and protected against water jets from any angle 4 bar IP65 Never dispose of the device in household waste...

Page 5: ...system The water quality on site is indicated by a flashing LED and can also be transmitted to an alert system via the floating contact An external analogue indicator can also be connected 3 Standard...

Page 6: ...coaxial power connector part no 102384 max 30V AC DC 2A 6 Connection for analogue output 4 20 mA max 24 V DC coaxial power connector part no 102384 max 24V DC _ 7 Connection for external power supply...

Page 7: ...service function 10 Insulation EPP 11 Connections on both sides 1 female 4 2 Dimensions 125mm 98mm 90mm 1 0 0 m m 4 3 Performance data PUROTAP i control 21 Display 3 LED green Good 0 to 200 S cm Disp...

Page 8: ...f pro tection IP 65 Power supply int Lithium battery 3 6 V 1200 mAh ER14250 Power supply ext Via micro USB 5 V DC 0 5 A max 5 Installation The PUROTAP i control 21 is installed in the water circuit fl...

Page 9: ...ctivity measurement correctly the maximum anticipated operating temperature of the system must be set PIN setting Max operating temperature PIN 1 PIN 2 PIN 3 25 C 0 OFF 0 OFF x any 20 C 0 OFF 1 ON x a...

Page 10: ...enables the measurement probe to be removed or replaced without any water leaking out and causing the pressure in the system to drop The fitted lock has a 1 female thread on both sides and causes virt...

Page 11: ...ld it down for more than 4 seconds and then release it All 3 LEDs should flash briefly once the device has now been recal ibrated 8 Spare parts Item no Designation 102 373 PUROTAP i control 21 replace...

Reviews: