21
9. Technical Data: MF-Serie
MF-Serie
Emmi®-MF30
Emmi®-MF60
External dimensions
L x W x H (mm)
265 x 165 x 225
330 x 175 x 290
Internal dimensions
L x W x H (mm)
238 x 138 x 100
302 x 152 x 150
Content
(liter)
3,0
6,0
Ultrasound power
(watt)
160/280
240/360
Heating power
(watt)
120
120
Generator
(Watt)
3 x 60
4 x 60
Outlet valve
yes
yes
Frequency
(kHz)
20/40
20/40
Power
(Watt)
280
360
Technical changes reserved.
End of the cleaning process
After ending of the cleaning process press ON/OFF to turn off the device. Hold down the key for 1
second. Remove the cleaned items in the stainless steel basket from the tube and rinse them with
clean water to solve any retarded dirt. The cleaning process can also be shortened by pressing
the «START / STOP» button slightly. The cleaning tub does not have to be emptied after each
cleaning process, unless the cleaning liquid is extremely soiled.
However, it is recommended that the device have to be emptied whenever it is not used for an
extended period of time. There are certain residues of soiling which can lead to corrosion in the
stainless steel tank. Do not use aqueous cleaning additives in the acidic range (e.g. fluoride, chlo-
ride or bromide ions). Explosive substances and flammable cleaning solutions cannot be used in
the stainless steel tub and should be sonicated.
After the connection cable has been pulled out of the device and the socket, the liquid can be
emptied over the tank corners or via the drain tab. It is recommended to dry the tub with a cloth.
Note:
Please ensure that a degassing time is imperative by hard water (high degree of lime).
In this case we recommend degassing the water for at least 2-3 min. before placing the
parts in to the device.
Summary of Contents for Emmi D Series