User Guide
–
Line Impedance Stabilization Network
13
6.0. Calibration
Before calibration, to abvoid electric shock or damage to the instruments, power off the
instrument.
LN2-100A Calibration Jig (CA-002-LISN)
Calibration Jig Connection
LISN’s calibration has to meet the impedance curve specified in CISPR 16. Because it is needed
to get the same result even in changed places or repeated tests.
LISN needs the exclusive jig for calibration.
Output BNC for measuring is terminated with 50
Ω
resistor and measure separately L1 and L2.
For the detailed impedance measurement, sweep measuring is needed. Calibration at a few
frequency may not show all the characteristic of LISN.