User Guide
–
Line Impedance Stabilization Network
9
3.0. Packaging
When
receive the products, please check the packing list and match it with actual items in the
package.
Please check the conditions of items; any damage, attachment, and looks is properly good.
Description
Quantity
Remark
Main body
1 set
Serial No.
Manual
1 ea
Calibration Jig
1 ea
CA-002-LISN
Table 1. Standard products package
# This packing list is based on standard package
Each purchased list shall be different according to the contract.
# Please contact if there are any missing parts.
4.0. Installation
Any use of the LISN without proper earth connection is strictly
Warning
prohibited. Due to the high leakage current to ground inherent to
this type of equipment, it is necessary to install a supplement
protective earthing wire from the protective earth terminal on the
front(rear) panel of the LISN to an appropriate earthing point of the
power mains.
If the main line is equipped with the fault current protection switch,
recommended to use an isolation transformer between main outlet and
the LISNs.
4.1. Protective earth connection
LISN is provided with earth connectors in the rear panel. The ground connection
between the LISN and the ground plate must be as short as possible and no inductive. The
earth point should be determined by an electrician authorized to do such work by
appropriate code or law.