background image

10 

PORTO

2

VENT™ CPAP

OS 

System Oper

ator’s Manual 

Part Number 1900-140 Rev I 

 

 

Pressure  Balanced  Patient  Valve  and  Inspiratory  Check  Valve    The  inspiratory  valve  opens  to  supply  flow  to  a 
spontaneous  breathing  patient  during  the  inspiratory  phase  of  breath  delivery.    The  expiratory  valve  is  pressure 
balanced  to  precisely  maintain  CPAP  levels  and  reduce  patient  effort  to  open  and  exhale  through  the  valve.   The 
patient valve gas collector port rotates 360

and is a 22 mm I.D. and 30 mm O.D. port to accept filters and spirometry 

devices. 

 

CAUTION:   

Do not use any other make or manufacturer of the pressure balanced patient valve that is not labeled 

for use with the  CPAP

OS 

device.   The patient breathing circuit must be connected as illustrated  in section  2c and 

described  in  section  3b  of  this  manual.    Failure  to  do  so  could  result  in  device  malfunction.    The  patient  and 
equipment must be observed constantly. 

 

Over Pressure Relief Valve    In the event of an occlusion and/or pressure build up in the CPAP

OS 

system, the Over 

Pressure Relief Valve is preset to relieve pressure at 22 cmH

2

O. 

 

Anti-asphyxiation Port   In the event of complete gas supply failure the Anti-asphyxiation Port allows the 
spontaneous breathing patient access to ambient air. 

 

Outlet Port Connector    The Outlet Port Connector is a locking bayonet connector used to attach the corrugated 
main tube, and engage the airway pressure line. 

Summary of Contents for Porto2vent CPAP os

Page 1: ...1 PORTO2VENT CPAPOS System Operator s Manual Part Number 1900 140 Rev I PORTO2VENT CPAPOS Operator s Manual...

Page 2: ...PORTO2VENT CPAPOS System Operator s Manual 2 Part Number 1900 140 Rev I...

Page 3: ...the CPAP Level 1b General Description 3b Applying the Breathing Circuit and Mask 1c Intended Use 3c Getting Started 1d Labels and Symbols 3d Monitoring the Patient 1e How CPAPOS Works 1f Breathing on...

Page 4: ...resentative will be able to assist you You may also contact the Emergent Respiratory LLC Customer Care and Service Assistance Line Toll Free at 866 411 4368 or e mail customercare eresp com 1 Introduc...

Page 5: ...erator adjustable levels throughout the breathing cycle independent of the patient s flow rate The System conserves the gas supply by terminating gas flow during the expiratory phase of the breath The...

Page 6: ...rehospital EMS environments CAUTION The CPAPOS unit and breathing circuit are intended only for use with a positive pressure face mask and should not be connected to endotracheal or oropharyngeal tube...

Page 7: ...7 PORTO2VENT CPAPOS System Operator s Manual Part Number 1900 140 Rev I 1d Labels and Symbols DISS Oxygen Inlet CPAP Adjustment Knob Increase clockwise Decrease counterclockwise Airway Pressure Gauge...

Page 8: ...tenance is recommended every two 2 years 1 Attach gas source to the GAS inlet located on the side of the device 2 Connect the CPAPos Breathing Circuit 3 Instruct Patient and Apply Mask 4 Set CPAP leve...

Page 9: ...zero liters per minute lpm 0 lpm To initiate flow of gas the patient must trigger the CPAP pressure generator DISS Oxygen Inlet Port The gas inlet port is a 9 16 standard DISS oxygen connector The spe...

Page 10: ...ure balanced patient valve that is not labeled for use with the CPAPOS device The patient breathing circuit must be connected as illustrated in section 2c and described in section 3b of this manual Fa...

Page 11: ...for such problems Bacteria Viral Filter The bacteria viral filter is connected to the Airway Pressure Line and outlet connector Corrugated Breathing Tube The corrugated main tube directs gas flow from...

Page 12: ...00 140 Rev I DISS Oxygen Inlet Port Over Pressure Relief Valve Internal Pressure Reducer Airway Pressure Gauge Anti asphyxiation Port Outlet Port Connector Bacterial Viral Filter CPAP Adjustment Regul...

Page 13: ...nt Knob clockwise until the Airway Pressure Gauge indicates approximately 5 cmH2O Unblock the outlet The unit should fully flow Block the outlet being careful to not block the Airway Pressure Line sma...

Page 14: ...rd make sure the oxygen connector is free from oil NOTE To turn the CPAPOS unit off disconnect the O2 supply hose from the source or turn off at the O2 cylinder CAUTION Certain data on the use of high...

Page 15: ...eathing Circuit contains the following components Six 6 foot corrugated main tube Area Ratio Pressure Balanced Exhalation Valve Bacterial Viral Filter Positive Pressure Medium Face Mask Small and Larg...

Page 16: ...ation Valve and Components Patient Valve Cap Pressure Balanced Exhalation Valve Patient Mask Connection 22 mm O D Exhalation Port with 3600 rotation and 22 mm I D 30 mm O D 22 mm Main Tube Connection...

Page 17: ...serve the needle indicator on the airway pressure gauge 3b Applying the Breathing Circuit and Mask Assembly Instructions The CPAPOS Breathing Circuit is pre assembled at the factory To attach the brea...

Page 18: ...t be connected as illustrated in section 2b of this manual Failure to do so could result in device malfunction The patient and equipment must be observed constantly CAUTION Do not clean or reuse the s...

Page 19: ...rning the CPAP adjustment knob clockwise and observe the needle indicator on the airway pressure gauge 3d Monitoring the Patient During operation monitor the following periodically 1 Monitor to assure...

Page 20: ...ure variation during CPAP treatment can cause fatigue and respiratory failure WARNING Potential side effects of continuous positive airway pressure may include fluid retention pneumothorax decreased c...

Page 21: ...ervice and repair center To return a CPAPOS unit call 1 866 411 4368 to obtain a Return Goods Authorization RGA number Warranty Emergent Respiratory LLC warrants the CPAPOS unit and its accessories to...

Page 22: ...40 60 psi 2 7 to 4 1 bar Operating Temperatures 00 F to 1200 F 18 to 500 C Storage Temperatures 400 F to 1600 F 40 to 700 C Over Pressure Relief 22 cmH2O max 35 cmH2O 100 lpm Anti asphyxiation Valve 8...

Page 23: ...Part Number 1900 140 Rev I This page intentionally left blank PORTO2VENT CPAP08 System Operator s Manual 23...

Page 24: ...Part Number 1900 140 Rev I 2731 Loker Avenue West Carlsbad CA 92010 For warranty registration of your CPAPOS unit visit www eresp com PORTO2VENT CPAPOS System Operator s Manual 24...

Reviews: