CPXH Temperature Calibration
The CPXH unit temperature measurement is factory calibrated. Use the following
instructions to perform periodic calibrations:
6.5.9
Draining Your Unit
Models 1800 and 2800 do not have a drain. To empty, use the indented side of the rim to
pour the used solution into a waste disposal unit, rinse the tank thoroughly and refill with
new solution.
Table 6.22 CPXH temperature calibration
Step
Action
1
The ultrasonic bath liquid may be at room temperature or may be
heated-up to a desired operating temperature (e.g. 40° C).
2
Press the Fn key 4 times until the Current Temperature icon starts
blinking. Press the Up key to select °F. Press the Fn key 2 more times to
return to the Ready State.
3
Press the On/Standby key to turn off the unit.
4
Simultaneously, press both the On/Standby and the Fn key. Only the
bottom left digits and the Current Temperature icon should turn on.
5
Stir the solution for 15 seconds to ensure thermal uniformity.
6
Wait 2 minutes after turning the unit on before taking measurements.
This allows for the display to be properly updated.
7
Use the Up/Down keys to change the display temperature to match the
actual tank temperature.
8
Press the On/Standby key to end calibration.
WARNING
General Warning
Do not immerse the unit in water. Unplug the unit from the power
source.
Summary of Contents for Branson CPX1800
Page 4: ...iv 100 214 294 REV 05 ...
Page 6: ...vi 100 214 294 REV 05 ...
Page 8: ...viii 100 214 294 REV 05 ...
Page 21: ...100 214 294 REV 05 11 Chapter 3 Delivery and Handling 3 1 Unpacking Your Unit 12 ...
Page 35: ...100 214 294 REV 05 27 Chapter 5 Installation and Setup 5 1 Installing Your Unit 28 ...
Page 76: ...72 100 214 294 REV 05 ...