Operating instructions
cod. 1592
cod. 1592
cod. 1592
cod. 1592017400
017400
017400
017400
1592017400 XLH260 GB r1.0 15.04.2005.doc
XLH260
13/15
13.
ALARM SIGNALLING
Message Cause
Outputs
“P1”
Thermostat probe failure
Compressor and heating outputs off
“P2”
Evaporator probe failure
Defrost and by time
“P3”
Humidity probe failure
Humidity regulation off
“HA”
High temperature alarm
Outputs unchanged
“LA”
Low temperature alarm
Outputs unchanged
“HHA”
High humidity alarm
Outputs unchanged
“HLA”
Low humidity alarm
Outputs unchanged
“dA”
Door switch alarm
Outputs depending on the odC parameter
“EAL”
External alarm
Other outputs unchanged
“BAL”
Serious external alarm
Outputs OFF
“PAL”
Pressure switch alarm
Outputs OFF
The alarm message is displayed until the alarm condition recoveries.
All the alarm messages are showed alternating with the room temperature except for the “
P1
” which is flashing. To reset the “EE” alarm and restart
the normal functioning press any key, the “
rSt
” message is displayed for about 3s.
13.1
SILENCING BUZZER
Once the alarm signal is detected the buzzer, if present, can be silenced by pressing any key.
13.2
ALARM RECOVERY
Probe alarms
: “
P1
” (probe1 faulty), “
P2
”, “
P3
”; they automatically stop 10s after the probe restarts normal operation. Check connections before
replacing the probe.
Temperature alarms
“
HA
” and “
LA
” automatically stop as soon as the thermostat temperature returns to normal values or when the defrost starts.
Humidity alarms
“
HHA
” and “
LHA
” automatically stop as soon as the humidity returns to normal values.
Door switch alarm
“
dA
” stop as soon as the door is closed.
External alarms
“
EAL
”, “
BAL
” stop as soon as the external digital input is disabled
Pressure switch alarm
“
PAL” alarm is recovered by switching OFF the instrument
.
10.
TECHNICAL DATA
Housing:
self extinguishing ABS;
Case:
frontal 210x230 mm; depth 87mm;
Mounting:
See par. 9;
Protection:
IP65
Connections:
Screw terminal block
≤
2,5 mm
2
wiring.
Power supply:
230Vac 50/60Hz
±
10% or 110Vac 50/60Hz
±
10%;
Power absorption: 10
VA max.
Display
:3 digits, red LED, 30.5 mm high; 3 digits, yellow LED low.
Inputs
: 2 NTC probes, one 4÷20mA input
Digital inputs :
door switch and configurable, free voltage. Max. distance 10m
Relay outputs:
compressor:
relay SPST 20(8) A, 250Vac;
defrost:
relay SPDT 16(3) A, 250Vac;
fans:
relay SPST 8(3) A, 250Vac
;
heater control:
relay SPST 20(8) A, 250Vac
;
Humidifying:
relay SPST 8(3) A, 250Vac;
Dehumidifying / Light:
relay SPST 16(3) A, 250Vac
Other output :
Alarm buzzer
(Standard)
Direct RS485
(optional)
Data storing
: on the non-volatile memory (EEPROM).
Kind of action:
1B.;
Pollution grade:
normal;
Software class:
A.
Operating temperature:
0÷60 °C.;
Storage temperature:
-25÷60 °C.
Relative humidity:
20
÷
85% (no condensing)
Measuring and regulation range: NTC probe:
-40÷110°C (-58÷230°F);
Resolution:
0,1 °C or 1°C or 1 °F (selectable).
Accuracy (ambient temp. 25°C)
: ±0,5 °C ±1 digit