Enable Advanced Diagnostics application (alternative)
If your CSI 9420 is not installed on a wireless network, you can perform the upgrade using
either a HART modem or a 375 or 475 Field Communicator.
WARNING!
The hazardous area rating available with the CSI 9420 does not permit either of the following
operations to be performed in a hazardous area. Do NOT open the device and connect to the
wired HART terminals in a hazardous area without taking the appropriate safety precautions
required by local, national, or international regulations.
Note
Connecting directly to the wired HART terminals on the CSI 9420 temporarily takes the device off of
the wireless network. If in range, it automatically rejoins the wireless network after the wired
connection is removed.
Method 1 - Using a wired HART modem
1.
Launch AMS Device Manager.
2.
Connect the CSI 9420 to an AMS Device Manager PC directly using a HART modem.
3.
Follow the steps in
Enable Advanced Diagnostics application (standard)
Method 2 - Using a 375 or 475 Field Communicator
1.
Use the lead set to connect the Field Communicator to the CSI 9420 terminal block.
2.
Power on the Field Communicator, and select
HART Application
from the main menu.
Depending on the Device Descriptor (DD) file in your CSI 9420, you may get a
warning message. Click
CONT
to proceed to the main menu.
3.
Select
Configure
or press 2 on the keypad.
4.
Select
Manual Setup
or press 2 on the keypad.
5.
Select
License
or press 6 on the keypad.
6.
Select
Configure License
or press 2 on the keypad.
7.
Select
Yes
or press 1 on the keypad.
This displays the serial number and request number. Call or email Product Support
and provide this information. Product Support will issue a registration key.
8.
Enter the registration key in the space provided and press
ENTER
.
2.4.2
CSI 9420 Data Collection: Overview
Data collection on the CSI 9420 includes taking an acquisition and storing it in memory
where it is available to be transmitted. AMS Machinery Manager obtains data from a CSI
9420 through the Data Import Server communication to the gateway device. You can view
or change data collection settings through AMS Machinery Manager, in the Data Import
program. You can set up policies and fine-tune your data collection based on time or
alerts.
Configuration
62
MHM-97408, Rev 15
Summary of Contents for CSI 9420
Page 4: ......
Page 10: ...Revision numbers in AMS Device Manager Figure 1 2 Introduction 4 MHM 97408 Rev 15 ...
Page 60: ...PeakVue spectrum Figure 2 35 Configuration 54 MHM 97408 Rev 15 ...
Page 92: ...Setup 86 MHM 97408 Rev 15 ...
Page 134: ...Product certifications 128 MHM 97408 Rev 15 ...