Gating configuration parameter State gated, active low
Figure 4-15:
13.
Click the Alert Setup tab (from the main Configuration page).
Configure Flow Analysis Alert
Figure 4-16:
14.
Click the Flow Analysis tab to select Configure Reverse Flow Detection, if desired.
The default setting is Disabled. Click the Disabled button to send the feature
command to the meter. Check for a response error. Click the Enable button if no
error response is received.
a. Enter the minimum reverse flow velocity above which to accumulate flow in the
reverse direction for this alert. Enter a positive value for the Reverse Flow Zero
Cutoff. Click the Next button to write the values to the meter. Check for an error
response. If no error response is received, click the Next button. The Detect
Reverse Flow enabled page displays. Click the Next button to display Detect
Reverse Flow disabled.
b. Click the Next button to display the Method Complete page if an error message
is returned.
Configuration
104
Gas Ultrasonic Flow Meter