Note
Before writing configuration changes to your meter, make sure you have saved the Configuration file
and Maintenance log.
Procedure
1.
Disable the Write Protect switch in the CPU Module to write any of the following
configuration parameters to your meter.
2.
Click the Setup Units tab to configure the system units (U.S. Customary or Metric
units), Volume units, Flow rate time units, Velocity units, Pressure units and
Temperature units.
a. Click Apply to write the parameters to the meter.
3.
Click the Setup Outputs tab to configure the Device Variables Mapping, Units,
Frequency/Digital outputs, Frequency and Digital Outputs 1 and 2, Analog outputs,
Digital Input, Pressure and Temperature.
a. Analog output 1 (HART) - Content (Primary Variable) displays Uncorrected Flow
Rate and is a read only attribute). Configure Direction (flow), Lower Range value,
Upper range value and Alarm Action and view the HART Parameters Tag, Date,
Descriptor, Message, Final Assembly Number Poll Address, Number of Response
Preambles.
b. Analog Output 2 - Content (Secondary Variable) displays Uncorrected Flow Rate
and has a read only attribute. Configure Direction (flow), Lower Range value,
Upper range value and Alarm Action. Map the Third and Four variables using the
Manual Setup wizard. Selections include Uncorrected Volume Flow Rate,
Pressure and Temperature.
4.
Click Apply to write the parameters to the meter after all of the data shown below is
entered.
a. Click the Frequency/Digital Outputs tab to configure Frequency/Digital Output
1, 2 and 3 Source and drive Mode. Select the Source for each Frequency/Digital
output and select the desired drive Mode. The Mode options are Open Collector
which requires an external excitation voltage and pull-up resistor or TTL mode
which outputs a 0-5 VDC signal (each Frequency output has an A and B output
phase). (Refresh Note: If changes are made to any Source variable on this page,
apply the changes and navigate to the Guided Setup page. Navigate back to the
Manual Setup for the changes to be reflected in other Manual Setup pages).
b. Click the Frequency and Digital Output 1 tab to configure the Content, (flow)
Direction, Channel B Phase frequency output, Lag forward, Lead Reverse or Lead
Forward, Lag Reverse (Phase B lags Phase A while reporting forward flow and lead
Phase A while reporting reverse flow or the opposite), Digital Output 1 Channel A
Content and Polarity, Channel B Content and Polarity, Maximum Frequency, and
Lower and Upper Range Units of Measure.
c. Click the Frequency and Digital Output 2 tab and repeat Step 3b to configure
Frequency and Digital Output 2 parameters.
Configuration
Installation manual
99