Emerson FB1100 Flow Computer Instruction Manual
D301752X012
August 2020
Installation
25
Figure 2-18: Ground Lug
1
Ground lug
DANGER
EXPLOSION HAZARD: Ensure the area in which you perform this operation is non-hazardous.
Performing this operation in a hazardous area could result in an explosion.
Once you have installed the unit, run a ground wire between the ground lug and a known
good earth ground. You route the ground wire through one of the conduit fittings.
Use stranded copper wire to earth ground and keep the length as short as possible.
Clamp or braze the ground wire to the ground bed conductor (typically a stranded copper
AWG 0000 cable installed vertically or horizontally).
Run the ground wire so that any routing bend in the cable has a minimum radius of 30.5 cm
(12 inches) below ground and 20.4 cm (8 inches) above ground.
For more information on grounding or if your installation uses cathodic protection, refer to
Site Considerations for Equipment Installation, Grounding, and Wiring
(D301452X012).
2.9
Terminal Plate
The terminal plate includes the various terminal blocks (TB) for power and I/O connections.
Note
AI/AO and DI connections are disabled in the FB1100 Flow Computer.