MPS-4009 Multi-Chemistry Power Supply Operations Manual Supplement, Doc. No. L01-497
43
InterMetro Industries Corporation
LiFe Mid-Grade (CALB) Battery Assembly Procedure
1. Align CALB battery with power supply enclosure and secure battery connector as shown
in figure below.
Align CALB Battery
2. Firmly grip the cable plug between thumb and index finger as shown below. Press the
tab on the battery terminal and insert it.
Note:
Secure the battery connector cable as well.
Plug Battery Connectors
3. Install the battery tray inside the Green Cubes Power Supply compartment as shown in
figure below.
Installation of Batteries