Oxymitter 5000
9-8
Instruction Manual
IM-106-350, Rev 2.2
July 2008
OXYMITTER 5000
REPAIR
Each of the following procedures details how to remove and replace a specific
component of the Oxymitter 5000.
Removal and
Replacement of Probe
1. Remove.
a. Turn off power to the system.
b. Shut off the calibration gases at the cylinders and the instrument air.
c. Disconnect the calibration gas and instrument air lines from the
Oxymitter 5000.
d. Remove left housing cover (31, Figure 9-3 or Figure 9-4).
e. Remove all signal and power wiring to the probe.
f. Remove insulation to access the mounting bolts.
g. Unbolt the Oxymitter 5000 from the stack and take it to a clean work
area.
h. Allow the unit to cool to a comfortable working temperature.
2. Replace.
a. Bolt the Oxymitter 5000 to the stack and install the insulation.
b. Connect all signal and power leads at the probe. Refer to Section 2:
Installation, for detailed wiring instructions.
c. Connect the calibration gas and instrument air lines to probe.
d. Install left housing cover (31, Figure 9-3 or Figure 9-4).
e. Turn on instrument air.
f. Restore power to the system; refer to "Power Up" in Section 5:
Startup and Operation of Oxymitter 5000 with Membrane Keypad or
"Power Up" in Section 6: Startup and Operation of Oxymitter 5000
with LOI. When the probe is at operating temperature, calibrate the
probe per "Calibration with Keypad".
NOTE
Recalibration is required whenever electronic cards or sensing cell is
replaced.
It is recommended that the Oxymitter 5000 be removed from the stack for all service
activities. The unit should be allowed to cool and be taken to a clean work area. Failure to
comply may cause severe burns.
Disconnect and lock out power before working on any electrical components. There is
voltage up to 115 VAC.
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...