Instruction Manual
IM-106-350, Rev 2.2
July 2008
1-15
Oxymitter 5000
SPECIFICATIONS
Oxymitter Specifications
O
2
Range
Standard
0 to 10% O
2
,
0 to 25% O
2
,
0 to 40% O
2
(via
FOUNDATION fieldbus)
Accuracy
±0.75% of reading or 0.05% O
2
, whichever is greater
System Response to Calibration
Gas
Initial – less than 3 seconds, T90 – less than 8
seconds
Temperature Limits
Process
32° to 1300°F (0° to 704°C) up to 2400°F (1300°C)
with optional accessories
Electronics Housing
-40° to 158°F (-40° to 70°C) amb ient
Electronics Package
-40° to 185°F (-40° to 85°C) [Operating temperature
of electronics inside of instrument housing, as read by
Foundaiton fieldbus.]
Local Operator Interface
-40° to 158°F (-40° to 70° C), [above 158°F (70°C) the
infrared keypad will cease to function, but the
Oxymitter 5000 will continue to operate properly.]
Probe Lengths
18 in. (457 mm)
12 ft (3,66 m)
3 ft (0,91 m)
15 ft (4,57 m)
6 ft (1,83 m)
18 ft (5,49 m)
9 ft (2,74 m)
Mounting and Mounting Position
Vertical or horizontal;
a spool piece, (P/N 3D39761G02), is available to
offset transmitter housing from hot ductwork.
Materials
Probe
Wetted or welded parts - 316L stainless steel (SS)
Non-wetted parts - 304 SS, low-copper aluminum
Electronics Enclosure
Low-copper aluminum
Calibration
Manual, semi-automatic, or automatic
Calibration Gas Mixtures
Recommended
0.4% O
2
, Balance N
2
8% O
2
, Balance N
2
Calibration Gas Flow
2.5 l/min (5 scfh)
Reference Air
1 l/min (2 scfh), clean, dry, instrument-quality air
(20.95% O
2
), regulated to 34 kPa (5 psi)
Electronics
NEMA 4X, IP66 with fitting and pipe on reference
exhaust port to clear dry atmosphere
Electric Noise
EN 61326-1, Class A
Certifications
General Purpose
Line Voltage
90-250 VAC, 48/62 Hz. No configuration necessary.
3/4 in. -14 NPT conduit port
Table continued on next page
APPROVED
US
C
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...