20
MODEL 1054B pH/ORP
SECTION 4.0
CONFIGURATION
TABLE 4-1. Configuration Work Sheet (pH)
Use this work sheet to assist in the configuration of the analyzer. Date: ____________________
RANGE
FACTORY SET
USER SET
A. Alarm 1 Setup (
AL1
)
1. Alarm Configuration (
On
/
OFF
)
On
_________
2. High or Low (
H-L
) (
Hi
/
Lo
)
Lo
_________
3. Hysteresis (
HYS
)
0-2.0 pH
0 pH
_________
4. Delay Time On (
don
)
0-255 sec.
000 Seconds
_________
5. Delay Time Off (
doF
)
0-255 sec.
000 Seconds
_________
B. Alarm 2 Setup (
AL2
)
1. Alarm Configuration (
On
/
FLt
/
OFF
)
On
_________
2. High or Low (
H-L
) (
Hi
/
Lo
)
Hi
_________
3. Hysteresis (
HYS
)
0-2.0 pH
0 pH
_________
4. Delay Time On (
don
)
0-255 sec
000 Seconds
_________
5. Delay Time Off (
doF
)
0-255 sec
000 Seconds
_________
C. Interval Timer (
IInntt
)
1. Active Status (
tOn
) (
oFF
/
on
)
oFF
_________
2. Interval Time (
int
)
Minimum 10 Minutes
10 Seconds
_________
3. Count (
cnt
)
1 to 60
5
_________
4. On Time (
ont
)
0 to 299 sec
1 Second
_________
5. Off Time (
OFt
)
0 to 299 sec
1 Second
_________
6. Duration (
dur
)
0 to 299 sec
2 Seconds
_________
D. Temperature Setup (
t-C
)
1. Display Temperature (
d-t
) (
oC
/
oF
)
*C
_________
2. Automatic Temperature
Compensation (
Atc
) (
on
/
oFF
)
on
_________
a. Manual Temp. Value
-10°C to 125°C
_________
E. Current Output Setup (
OUt
)
1. mA Output (
CUr
) (
020
/
420
)
420
_________
2. Display Current Output (
d-O
) (
Pct
/
doc
)
doc
_________
3. Dampen Current Output (
dPn
)
0-255 sec.
0.0 Seconds
_________
F. Electrode Diagnostics Setup (
InP
)
(
on/oFF
)
oFF
1. Temp compensated impedance (
rES
)
2. Impedance increase before calibration (
CAL
)
20-200%
20%
_________
3. Cracked glass impedance low limit (
SHt
)
5-600 megohms
20 megohms
_________
4. Aged electrode impedance high limit (
OId
)
200-1999 megohms
1000 megohms
_________
G. Default Setup (
dFt
)
1. Relay 1 Default (
rL1
) (
non
/
oFF
/
on
)
non
_________
2. Relay 2 Default (
rL2
) (
non
/
oFF
/
on
)
non
_________
3. Current Output Default (
Cur
) (
non
/
cur
)
non
_________
H. Automatic Buffer Mode (
AbF
)
1. Configuration (
on/off
)
on
_________
I. Keyboard Security Setup (
COd
)
1. Keyboard Security Required
001-999
_
_________
2. Keyboard Security Not Required
000
000
_________
Alarm Setpoints
1. Alarm 1 (
SP 1
)
0-14 pH
0.00 pH
_________
2. Alarm 2 (
SP 2
)
0-14 pH
14.00 pH
_________
Current Output
1. Zero (0 or 4 mA) (
LO
)
0-14 pH
0.00 pH
_________
2. F.S. (20 mA) (
HI
)
0-14 pH
14.00 pH
_________