127
July 2017
Reference Manual
00809-0100-2230, Rev CA
Table C-2.
FD_SIMULATE elements
79
FAILED_ALM
Alarm indicating a failure within a device which makes the device non-operational.
80
MAINT_ALM
Alarm indicating the device needs maintenance soon. If the condition is ignored,
the device will eventually fail.
81
ADVISE_ALM
Alarm indicating advisory alarms. These conditions do not have a direct impact on
the process or device integrity.
82
FAILED_ENABLE
Enabled FAILED_ALM alarm conditions. Corresponds bit for bit to the
FAILED_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected.
This parameter is the Read Only copy of FD_FAIL_MAP.
83
FAILED_MASK
Mask of FAILED_ALM. Corresponds bit of bit to FAILED_ACTIVE. A bit on means that
the condition is masked out from alarming.
This parameter is the Read Only copy of FD_FAIL_MASK.
84
FAILED_ACTIVE
Enumerated list of failure conditions within a device. All open bits are free to be
used as appropriate for each specific device.
This parameter is the Read Only copy of FD_FAIL_ACTIVE.
85
MAINT_PRI
Designates the alarming priority of the MAINT_ALM
86
MAINT_ENABLE
Enabled MAINT_ALM alarm conditions. Corresponds bit for bit to the
MAINT_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected.
This parameter is the Read Only copy of FD_OFFSPEC_MAP
87
MAINT_MASK
Mask of MAINT_ALM. Corresponds bit of bit to MAINT_ACTIVE. A bit on means that
the condition is masked out from alarming.
This parameter is the Read Only copy of FD_OFFSPEC_MASK.
88
MAINT_ACTIVE
Enumerated list of maintenance conditions within a device.
This parameter is the Read Only copy of FD_OFFSPEC_ACTIVE.
89
ADVISE_PRI
Designates the alarming priority of the ADVISE_ALM
90
ADVISE_ENABLE
Enabled ADVISE_ALM alarm conditions. Corresponds bit for bit to the
ADVISE_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected.
This parameter is the Read Only copy of FD_MAINT_MASK & FD_CHECK_MASK.
91
ADVISE_MASK
Mask of ADVISE_ALM. Corresponds bit by bit to ADVISE_ACTIVE. A bit on means
that the condition is masked out from alarming. This parameter is the Read Only
copy of FD_MAINT_MASK & FD_CHECK_MASK.
92
ADVISE_ACTIVE
Enumerated list of advisory conditions within a device. All open bits are free to be
used as appropriate for each specific device.
This parameter is the Read Only copy of FD_MAINT_ACTIVE & FD_CHECK_ACTIVE.
1
Diagnostic Simulate
Value
Bit string
4
Writable. Used for diagnostics when simulation is
enabled
2
Diagnostic Value
Bit string
4
Current diagnostics detected by the device.
3
Enable
Unsigned 8
1
Enable/Disable simulation. Dynamic, so simulation
will always be disabled after a device restart.
Index
Number
Parameter
Description
Summary of Contents for Rosemount 2230
Page 1: ...Reference Manual 00809 0100 2230 Rev CA July 2017 Rosemount 2230 Graphical Field Display ...
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2230 Rev CA Title Page July 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2230 Rev CA Introduction July 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2230 Rev CA Overview July 2017 Overview ...
Page 44: ...34 Reference Manual 00809 0100 2230 Rev CA Installation July 2017 Installation ...
Page 154: ...144 Reference Manual 00809 0100 2230 Rev CA Index July 2017 Index ...