115
Reference Manual
00809-0100-4570, Rev DA
Operation
November 2017
Operation
Figure 7-76. Firmware Update
Communication quality test
The communication quality test tool tests the connection quality between the computer and the
selected device(s). If packets not answered, or received with errors, there is an indication in the results
log.
In cases where the quality is not 100 percent, it is recommended that the RS-485 cable is to be checked
for existence of proper resistors on both sides and verified to be of shielded and twisted-pair type and
that the cable is designed for RS-485 communication protocol.
An example for a good test result: an average delay of 97msec (very low), no lost packets and no errors
received.
1. Set the number of cycles between 20 to 1000.
2. Select
Start Test
to begin the testing. The test will cycle through until terminated.
3. Select
Stop Test
to terminate the operation.
4. Select
Show Log
to open the results in a Notepad text file.
It is important to test the quality of the communication between the computer running the Rosemount
3DVision Server and the devices after initial setup and first connection.
Summary of Contents for Rosemount 5708
Page 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Page 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Page 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Page 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Page 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Page 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...