12. Configure the vessel full and empty calibration levels.
shows the full and empty
calibration fields to set.
It is possible to set the full and empty calibration as measured from the top (Distance) of the vessel or
measured from the bottom (Level) of the vessel.
Set the distance from the top of the vessel to set 100% (20 mA) (see
) in
the
Full Calibration
section.
OR
Set the maximal material level in the Level (Bottom) field. Dead band is 1.64 ft (0.5 m) from the
top of the Rosemount 5708 antenna case.
Set the
Empty Calibration
by setting the level or the distance from the top of the vessel to set the
0% (4mA) (see
). Default value is set to the bottom of the vessel.
Select the
Default
button to set the full and empty calibration levels to default values.
Select
View Table
to open a window showing a summary table of each Rosemount 5708 in the
vessel and the actual full and empty calibration levels configured, see
.
13. Select
Finish
.
Figure 7-12. Configuration Wizard Step 4 - Full and Empty Calibration Levels
Figure 7-13. Full and Empty Calibration Levels Summary Table
14. When the configuration wizard settings are completed, select the
Finish
button. The parameters will
be uploaded to all devices in the vessel.
Summary of Contents for Rosemount 5708
Page 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Page 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Page 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Page 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Page 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Page 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...