109
Reference Manual
00809-0100-4570, Rev DA
Operation
November 2017
Operation
7.5.6
Echo Curve Analysis
Performing Echo Curve Analysis
The Echo Curve Analysis tool collects information from the Rosemount 5708 and presents it in a chart
-type report for each scan.
To perform an Echo Curve Analysis:
1. On the
Device
menu, select
Echo Curve Analysis…
or press
Ctrl+Z
.
2. Select the device(s) from the tree view pane. The devices are organized in a Site-Vessel-device view.
3. Optional: Select
All Echoes Info
to perform an analysis on the Rosemount 5708 to verify behavior in
the vessel.
4. Select
Start
to initiate a single time Echo Curve analysis.
5.
Open Viewer
to open the Echo curve viewer.
If a single analysis scan is performed, then the Echo Curve Viewer will automatically pop up. To view
the echo curve, see next chapter.
Figure 7-71. Activate Echo Curve Analysis
To perform continuous echo curve analysis, set the
Every (mins)
and the
Duration (hours)
and select
Start
.
Every (mins)
: Sets the pause time between analysis scans, the pause time is measured in minutes and
by default set to five minutes. The pause time can be set between 3 to 1000 minutes.
Duration (hours)
: Set the period of time in hours to perform continuous analysis. Minimal value is 0.1
hour (six minutes) and maximal value is 999 hours (over 41 days).
Summary of Contents for Rosemount 5708
Page 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Page 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Page 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Page 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Page 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Page 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...