3. Attach the deflection plate to the bottom pipe by using two hose clamps. The
deflection plate allows you to measure down to the bottom of an empty tank. Make
sure that the lower pipe is cut to leave room for a deflection plate and about 20 mm
(0.8 in.) free space between the tank bottom and the deflection plate.
Note
7 m (23 ft) or longer still-pipes might require anchoring to better withstand tank
movements.
A. Pipe coupling
B. Hose clamps
C. Deflection plate
D. 60 mm
E. 20 mm