4.5.3
Surface Echo Tracking
The Surface Echo Tracking function can be used to eliminate problems with certain types
of ghost echoes below the product surface. This may, for example, occur in still-pipes as a
result of multiple reflections between the pipe wall, flange and antenna. In the tank
spectrum these echoes appear as amplitude peaks at various distances below the product
surface.
To activate this function, ensure that there are no disturbing echoes above the product
surface and select the
Always Track First Echo
check box.
To open the
Surface Echo Tracking
window:
Procedure
1. In the
TankMaster WinSetup
workspace, click the right mouse button on the desired
Rosemount 5900C icon.
2. Choose the
Properties
option from the popup menu.
3. In the
RLG Properties
window, select the
Advanced Configuration
tab.
4. Click the
Surface Echo Tracking
button:
Figure 4-9: The WinSetup Surface Echo Tracking Window
Echo Time Out
Use Echo Time Out to define the delay time until the gauge will start searching for a
surface echo after it has been lost. Until this period of time has elapsed, the gauge will not
start searching or trigger any alarms.
Configuration
Reference Manual
September 2019
00809-0100-5901
134
Reference Manual
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...