6.3.3
Error messages
shows a list of error messages that may appear on the integral display of the
Rosemount 2410 Tank Hub and in the Rosemount TankMaster program. You also have the
option to view Input register 1002 for an overview of active device errors.
For each error message that may appear, detailed information can be found in Input
registers 6000 to 6030 as shown in
.
Table 6-4: Error Messages for the Rosemount 5900C
Message
Description
Action
RAM error
Input register no. 6000.
A gauge data memory (RAM) error has
been detected during the startup
tests.
Note
This automatically resets the gauge.
Serious RAM problem:
Bit 0: DSP RAM
Bit 1: DSP stack
Bit 2: DSP RAM checksum
Bit 3: DSP RAM low
Contact Emerson Automation
Solutions /Rosemount Tank Gauging
service department
FPROM error
Input register no. 6002.
An error in the gauge program
memory (FPROM) has been detected
during the startup tests.
Note
This automatically resets the gauge.
Serious FPROM problem:
Bit 0: DSP Boot checksum
Bit 1: DSP Boot version
Bit 2: DSP Application checksum
Bit 3: DSP Application version
Bit 4: FPROM device
Bit 5: FPROM erase
Bit 6: FPROM write
Bit 7: FPROM active block not used
Service and troubleshooting
Reference Manual
September 2019
00809-0100-5901
232
Reference Manual
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...