Figure 4-11: SW1 switches on the backplane
Note
The GC can be connected (or remain connected) to the local network on Ethernet2
(
TB11
) on the backplane while the DHCP feature on Ethernet1 is being used.
This starts the GC’s DHCP server feature. The server typically takes approximately
20 seconds to initialize and start up.
3. Wait for 20 seconds and then do the following to ensure that the server has
provided an Internet protocol (IP) address to the PC:
a) From the PC, go to
Start
→
Control Panel
→
Network Connections
.
The
Network Connections
window lists all dial-up and local area network
(LAN)/high-speed Internet connections installed on the PC.
b) In the list of LAN / high speed Internet connections, find the icon that
corresponds to the PC-to-GC connection and check the status that displays
beneath the Local Area Connection.
It should show the status as
Connected
. The PC is now capable of connecting
to the GC.
If the status is
Disconnected
, it may be that the PC is not configured to
accept IP addresses; therefore, do the following:
4. Right-click the
Properties
icon.
The
Local Area Connection Properties
window displays.
5. Scroll to the bottom of the
Connection
list box and select
Internet Protocol (TCP/
IP)
.
6. To configure the PC to accept IP addresses issued from the GC, select the
Obtain an
IP address automatically
and
Obtain DNS server address automatically
check
boxes.
7. Click
OK
to save the changes and to close the
Internet Protocol (TCP/IP) Properties
window.
8. Click
OK
to close the
Local Area Connection Properties
window.
9. Return to the
Network Connections
window and confirm that the appropriate icon’s
status reads
Connected
.
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
65
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......