Postrequisites
If the icon still reads
Disconnected
Troubleshoot DHCP connectivity issues
.
NOTICE
If you power cycle the GC, you will lose connectivity.
4.4.8
Connect to the gas chromatograph (GC) using
Rosemount MON2020
To connect to the GC using the RJ45 Ethernet1 connection:
Procedure
1. Start Rosemount MON2020.
The
Connect to GC
window displays.
2. Locate the default
Direct-DHCP
under the
GC Name
column.
This GC directory is created automatically when Rosemount MON2020 is installed.
You can rename the GC, but do not change the IP address that it references,
192.168.135.100
.
3. Click the associated
Ethernet
button.
Rosemount MON2020 prompts you to enter a user name and password.
4. Enter your user name and password.
5. Rosemount MON2020 connects you to the GC.
Installation and start-up
Reference Manual
June 2022
2-3-9000-744
66
Emerson.com/Rosemount