Event log
•
Various analysis reports
1.4
Theory of operation
Related information
1.4.1
Thermal conductivity detector (TCD)
One of the detectors available on the gas chromatograph (GC) is a TCD, which consists of a
balanced bridge network with heat sensitive thermistors in each leg of the bridge. Each
thermistor is enclosed in a separate chamber of the detector block.
One thermistor is designated the reference element, and the other thermistor is
designated the measurement element. See
for a schematic diagram of the TCD.