5.13
Replace the Ingress Protection filter
The Ingress Protection (IP) filter (part number 00628-9000-0001) protects the sensor
inside the gas sensor module from ingress of fluids and solids. Replace the IP filter if it
becomes clogged with foreign matter.
CAUTION
Ingress Protection (IP) filter
If you don't install the IP filter, damage may occur to the sensor inside the gas sensor
module.
Do not operate the transmitter without the correct IP filter installed in the gas sensor
module.
When installing the IP filter, verify that the IP filter gasket is in place, is properly aligned,
and is not blocking the white filter media. Refer to
.
When handling the IP filter, avoid contact with the filter media.
Verify that all three legs are fully latched by pushing upward on each leg of the IP filter.
Avoid getting water inside the IP filter.
Do not attempt to clean the IP filter.
Do not rinse or spray the IP filter with water.
Do not immerse the IP filter in water.
Figure 5-1: IP Filter
A. IP filter housing
B. IP filter gasket
C. Filter media
Procedure
1. Remove the IP filter from the underside of the gas sensor module housing by
compressing the three snap tabs and pulling the IP filter downwards.
Reference Manual
Operation and maintenance
00809-0100-4928
September 2019
Rosemount 928
165
Summary of Contents for Rosemount 928
Page 1: ...Reference Manual 00809 0100 4928 Rev AE September 2019 Rosemount 928 Wireless Gas Monitor ...
Page 4: ...4 ...
Page 8: ...Contents Reference Manual September 2019 00809 0100 4928 viii Emerson com Rosemount ...
Page 106: ...Configuration Reference Manual September 2019 00809 0100 4928 106 Emerson com Rosemount ...
Page 178: ...Troubleshooting Reference Manual September 2019 00809 0100 4928 178 Emerson com Rosemount ...