Status
Description
4-20mA
output
Analog
output
Fault
relay
LED
indicator
Test rate
S92
Restore from
wrong voltage
1mA
0v
Open
4Hz
blinking
orange
After wrong
voltage
V81
1mA
0v
Open
4Hz
blinking
orange
Every 30msec
V82
1mA
0v
Open
4Hz
blinking
orange
Every 30msec
V83
Wrong vin
1mA
0v
Open
4Hz
blinking
orange
Every 30msec
P71
Program
memory CRC
failure
1mA
0v
Open
4Hz
blinking
orange
At start-up
P72
Faulty E2PROM
1mA
0v
Open
4Hz
blinking
orange
At start-up or
upon setup/
parameters
burning
P74
RAM CRC failure 1mA
0v
Open
4Hz
blinking
orange
Every 30msec
P75
None of the
parameters exist
1mA
0v
Open
4Hz
blinking
orange
At start-up
P76
RAM parameters
CRC failure
1mA
0v
Open
4Hz
blinking
orange
Every hour
F31
Sensor 4.5µ
circuit failure –
no signal
1mA
0v
Open
4Hz
blinking
orange
Auto/manual BIT
F32
Sensor 5µ circuit
failure – no
signal
1mA
0v
Open
4Hz
blinking
orange
Auto/manual BIT
F33
Sensor 4µ circuit
failure – no
signal
1mA
0v
Open
4Hz
blinking
orange
Auto/manual BIT
F34
Sensor 4.4µ
circuit failure –
no signal
1mA
0v
Open
4Hz
blinking
orange
Auto/manual BIT
F38
Wrong AFE
offset
1mA
0v
Open
4Hz
blinking
orange
Every 30msec