Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-1
Preventive Maintenance Schedule . . . . . . . . . . . . . . . . . . page 7-1
Control Unit Preventive Maintenance . . . . . . . . . . . . . . . . page 7-2
OVERVIEW
This section describes the routine maintenance of the Rosemount Analytical
OPM 4000 Opacity/Dust Density Monitor. The amount of maintenance
required depends on the environment to which the instrument is subjected. A
unit that is subjected to extremely dusty conditions, vibration, and movement
will require considerably more maintenance than a unit that is rigidly mounted
in a clean, dust-free atmosphere.
PREVENTIVE
MAINTENANCE
SCHEDULE
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
Item to Check
Frequency
Procedure
Zero/Span
Daily
Check that readings are within
specification (±2%). Refer to
Section 6: Zero and Span Calibration.
Fault conditions
Daily
Check for fault conditions; refer to
"Fault Messages" in Section 4:
Operation.
Projection lens
Monthly or as required
Clean with cloth and lens cleaning
fluid.
Beam alignment
Monthly or as required
Check alignment; refer to "Beam
Alignment Procedure" in Section 2:
Installation.
Transceiver and retro
reflector windows
Monthly or as required
Clean with cloth and lens cleaning
fluid. To access, refer to "Air Plenum"
in Section 2: Installation.
Zero reflector
Monthly or as required
Clean with optical cloth and lens
cleaning fluid. Refer to Section 5:
Zero Jig and Neutral Density Filters
for additional help.
All daily and monthly
checks
3 months
COM Audit
3 months
Perform COM Audit per EPA
regulation 40 CRF, 60 App. B, PS-1.
Air blower
3 months
Check for excessive noise.
Air filters
3 months
Replace.
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...