Instruction Manual
IM-105-4000, Rev 1.3
October 2008
4-3
OPM 4000
Alarm Set Point
Adjustments
1. Open the hinged cover and press the EDIT (ESC) button [the red Edit
(ESC) LED will light]. The message display will read QUICK MENU.
2. Press the ENTER key then the EDIT
▼
key 1 time until the Display
reads ALARM SET POINT.
3. Press the ENTER key; the Edit Value Light will flash.
4. Press and hold the EDIT
▲
or EDIT
▼
key until the desired set point in
% opacity is displayed, i.e. 20.00 is 20.00% opacity.
5. Press ENTER then press the EDIT key 4 times to return to normal
operation.
Alarm Time Delay
Adjustment
1. Open the hinged cover and press the EDIT (ESC) button [the red Edit
(ESC) LED will light]. The message display will read QUICK MENU.
2. Press the ENTER key then the EDIT
▼
key 2 times until the Display
reads ALARM TIME DELAY.
3. Press the ENTER key; the Edit Value Light will flash.
4. Press and hold the EDIT
▲
or EDIT
▼
key until the desired time in
seconds is displayed, i.e. 15 is 15 seconds delay.
5. Press ENTER then press the EDIT key 4 times to return to normal
operation.
4-20 mA Outputs
The OPM 4000 is equipped with two 4-20 mA outputs. The ranges are set
during final test to the information supplied by the customer.
Maximum current loop resistance is 800 ohms.
AOUT1 (instantaneous opacity, zero/span when in check mode) is connected
at terminals 11 and 12 and supplies the opacity signal to a recorder or DAS
and to the service module for display at the sensor location.
AOUT2 (average opacity, zero/span when in check mode) is connected at
terminals 13 and 14.
Recorder or Other
Device Calibration
On the control unit front panel press the CAL. AUTO/MAN key for a
maintenance zero and span cycle. Both outputs will correspond to the opacity
display value.
Auto Cal Timer
Adjustment
1. Open the hinged cover and press the EDIT (ESC) button [the red Edit
(ESC) LED will light]. The message display will read QUICK MENU.
2. Press the ENTER key then the EDIT
▼
key 3 times or scroll down until
the Display reads AUTOCAL CYCL TMR.
3. Press the ENTER key; the Edit Value Light will flash.
4. Press and hold the EDIT
▲
or EDIT
▼
key until the desired set point in
hours is displayed, i.e. 24.00 equals 24 hour intervals.
5. Press ENTER then press the EDIT key 4 times to return to normal
operation.
Remote Calibration
Cycle Initiate
To have an external device cause the OPM 4000 to begin calibration cycle,
connect an N.O. dry contact to terminal 8L and 9. A momentary closure will
initiate a 3-minute zero and a 3-minute span check cycle. Total time of
calibration check is 6 minutes.
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...