Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
5-2
Installation
The zero jig (Figure 5-2) on-line test reflector is inserted over the transceiver
lens. The filters provided are marked in single-pass opacity; this is the same
as double-pass with a correlation factor (OPLR) of 0.50.
1. Hold the zero jig in a position to see the reflector. Turn the fine
adjustment (large knob) so the iris begins to open, exposing more of the
reflector. This direction will cause the monitor to lower opacity. To raise
opacity, turn in the opposite direction so the iris begins to close.
2. Loosen clamps on the transceiver and swing it open.
3. Clean the lens using alcohol and a clean lint free paper that will leave no
film on the lens.
4. Place the zero jig on the two alignment pins (Figure 5-2) with the
reflector facing toward the transceiver (away from the stack).
5. Secure the device to the transceiver by tightening the mounting screw.
6. Using the fine adjustment (large knob) start by raising the opacity level
above zero then slowly turn the knob in the opposite direction to just
reach zero. If you cannot reach zero, or you cannot come above zero,
see "Coarse Adjustments".
7. Place the neutral density slides in the slot provided on the top of the
device just behind the mounting screw. The system should read the
same value written on the label within ±2%.
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...