Instruction Manual
IM-105-4000, Rev 1.3
October 2008
OPM 4000
7-4
CORRECTIVE
MAINTENANCE
NOTE
Make sure to carefully record the wiring connections and locations before
disconnecting. This will aid in the installation of the unit.
Lamp Replacement
1. Remove the transceiver cover by removing the screw below the target
viewing window and pulling the housing straight back until it clears the
optical plate.
2. Remove the lamp wire connector from the J2 socket on the R/M signal
processor board.
3. Remove the two socket head cap screws (Figure 7-1) using an Allen
wrench. Remove the lamp holding fixture
4. Turn the set screws to loosen the clamp for lamp removal.
5. Remove the lamp assembly consisting of the lamp, wires and
connector.
6. Install new lamp assembly in the lamp holding fixture and tighten set
screws.
7. Install lamp holding fixture and secure with the two socket head cap
screws. Connect J2.
8. Calibrate the OPM 4000 per Section 3: Startup and Calibration.
9. Install cover.
Figure 7-1. Lamp/Bulb Removal
Disconnect all power before installing or replacing components. Failure to disconnect power
may result in electrical shock and/or damage to the equipment.
38450029
Set
Screws
Socket Head
Cap Screw
Socket Head
Cap Screw
Lamp
Lamp Holding
Fixture
Summary of Contents for Rosemount Analytical OPM 4000
Page 2: ......
Page 6: ......
Page 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Page 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Page 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Page 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Page 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Page 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Page 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Page 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Page 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Page 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Page 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...