2.
Press
ENTER
.
T REF appears.
3.
If needed, change the reference temperature from the factory default 77 °F (25 °C)
to a different reference temperature for the process. Press
ENTER
.
UNIT appears.
4.
Press the up or down arrow to select the desired measurement units: uS
(microsiemens), mS (millisiemens), none (no units displayed), % (percent), or ppm
(parts per million); then press
ENTER
.
NUM PTS appears.
5.
Press up or down arrows to select the desired number of data points for a custom
conductivity curve. Selecting 2 will generate a linear relationship for conductivity
and concentration at the given reference temperature.
a. Enter the concentration for Pt. 1 (displayed as uS 1). Press
ENTER
.
b. Enter the known conductivity for Pt. 1 in uS/cm. Press
ENTER
.
c. Complete this process for additional known data points. Press
ENTER
.
CALC CUST appears briefly.
d. Press
ENTER
.
PROCESSING appears briefly; then APPLY CUST appears.
e. Press
ENTER
to register the custom curve into memory and return to the
SETUP
CUST
screen.
The custom curve will now be used to display and output all conductivity measurement
when CUST is selected in the display menu for measurement type.
6.5
Restore to default factory settings
6.5.1
Configuration overview
All user-defined configurations can be erased to return the transmitter to the factory
default settings.
6.5.2
Returning to factory default settings
1.
Press
PROG
on the remote controller.
2.
Press
NEXT
until DEFAULT appears in the display. Press
ENTER
.
3.
Use the up and down arrows to toggle between NO and YES. Press
ENTER
to return
to the factory default settings.
Programming - basics
Instruction Manual
37
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...