36
13. TABLE LIST FOR FULL CAPACITY
14. ERROR CODES
During the initial power-on testing or during operation it is possible the scale may show an
error message. The meaning of the error messages is described below.
If an error message is shown repeat the procedure that caused the message, turning the scale
on, calibration or other functions. If the error message still is shown then contact your dealer
for further support.
Spec
unit
3kg x 0.1g
6kg x 0.2g
15kg x 0.5g
30kg x 1g
g
3.0009
6.0018
15.0045
30.009
Ib
6.0018
15.0045
30.009
60.018
oz
105.045
210.09
520.18
1050.45
ERROR CODE
POSSIBLE CAUSES
HANDLING
E1
EPROM data lose.
Re-calibrate the scale
1.Load cell damaged.
1.Replace the load cell.
E2
2.Turn on the scale with
something on the pan.
2.Correct the operation.
E3
The sample weight for setting up
Percent weighing is much less than
the weight on the pan.
Correct the operation.
E4
The sample weight is less than
10d.
Increase the weight on the
pan.
E5
Wrong values are inputted.
(when Hi is less then LO)
Reset the limit values.
E6
The subtraction value is larger than
the total accumulated value
Correct the operation.