UM-155-290-Limpet User Manual
–
V2.0
6 |
P a g e
4.
Installing the software
•
Insert the EML USB stick received with the EML Limpet Logger (or download from the EML
website
–
).
•
Run t
he ‘.exe’ file contained on the
device or the download.
•
Follow the instructions to install the software.
•
After
installation, run the ‘EML Limpet
Logger GUI.exe’ to open the software
.
5.
Configuring the logger
Connect logger to a PC
•
Connect one end of the serial cable to the Limpet Logger and the other end to a free serial
USB port on the Computer.
•
Select ‘Auto detect’ on the Port combo box then Press the ‘Connect’ button. This will
automatically select the appropriate COM port and connect the logger GUI to it.
Run the ‘Configuration wizard’
Press the
‘Config
W
izard’
button. This guides the user through the main requirements for configuring
the logger. These are summarised below:
•
Clear logger memory
•
Set date-time
•
Select measurement unit
•
Select calibration factor
•
Select memory mode
Manually configure the logger
This section describes in more detail the steps carried out in the Configuration Wizard. For manual
configuration of the logger, it is recommended that the user carries out the following tasks in order:
Clear logger memory:
Press
‘Clear memory’
button on the main screen of the software and, if you are sure, press
‘Yes’
when
prompted.
This ensures that the new logger configuration makes use of the whole logger memory. Note that the
memory must be cleared before a new configuration is loaded. Therefore, if there are data on the
logger which are needed, exit the configuration wizard at this point, download the data, and then
return to the configuration wizard.
Set date-time:
Automatically set the date-time:
Press
‘Auto set date
-
time’
button on the main screen of the
software. This will set the logger’s time to the PC time.
Manually set the date-time:
Go to
‘Options’
then
‘Set Date and Time Manually’
. Use the drop-down
menus to set the logger’s time manually.