12
Relative Button
Repeatedly pressing the Relative button allows switching between
relative mode and auto mode.
Cannot be activated for some functions.
Hz/Duty Button
Repeatedly pressing the Hz/Duty button allows switching between
measuring frequency and duty cycle when the Hz/Duty function
is selected.
Operating Procedure
AC Voltage/DC Voltage Measurement
Connect the plug of the black (negative) tipped measuring conductor
into the COM socket and the plug of the red (positive) conductor for
measuring voltage into the INPUT socket. Turn the circular switch to
select the DC voltage function, marked as , or the AC voltage func-
tion, marked as
. Range will be automatically set to AUTO mode;
you can switch to manual range selection using the Range button.
Place the measuring tips onto to spot you want to measure. The
voltage value and polarity (for DC voltage) will appear on the screen.
If you do not know the voltage range beforehand, set the largest
possible range and gradually reduce it as you measure. Connect the
measuring tips to the device or circuit where you will be measuring
voltage. Turn on the device you want to measure. The voltage value
will appear on the screen.
If „OL“ appears on the screen while in manual mode, set a higher
range.
Note: To prevent electric shock and damaging of the device, do not
connect the multimeter to voltage higher than 600 V AC/DC CAT III
and 750 V AC/1 000 V DC CAT II.
Alternating Current/Direct Current Measurement
Note:
When measuring current up to 600 mA, connect the plug of the
black (negative) tipped measuring conductor into the COM socket
and the plug of the red (positive) conductor for measuring voltage
into the µA/mA socket. Switch to the function marked as
or
and select the range.
Summary of Contents for M0430
Page 2: ...2 1 2 3 9 4 8 5 10 7 6 1...
Page 114: ...114 CAT II IV M0430 1 000...
Page 115: ...115 30 V AC rms 42 60 DC...
Page 116: ...116 9 1604 6F22 AC DC AC DC...
Page 124: ...124 OL WH3303 DM633 10 AC 1 COM INPUT 2 3 1 1 1 2 1 1 COM INPUT Select 30...
Page 125: ...125 COM INPUT Select OL 1 1 M 2 OL COM INPUT Select OL 1 hFE 2 COM INPUT 2 3 NPN PNP...
Page 126: ...126 4 hFE 2 A B 1 Select C F 2 COM K INPUT 3 K 20 C 300 C 300 C 1 COM INPUT 2 3...
Page 181: ...181 BG M0430 EN 61010 1 CAT III 600 V CAT II 1 000 V 2 CAT III CAT II IV M0430...
Page 182: ...182 1 000 V...
Page 183: ...183 30 V 42 V 60 V...
Page 184: ...184 9 V 6F22 1604 C F 1 M0430 5...
Page 191: ...191 Relative Relative Hz Duty Hz Duty Hz Duty COM INPUT AUTO Range OL 600 V...
Page 193: ...193 1 1 A 1 mV 2 1 A 1 mV COM INPUT Select 30 COM INPUT Select OL 1 1 M...
Page 195: ...195 3 K 20 C 300 C 300 C 1 COM INPUT 2 3 600 F 10...
Page 196: ...196 1 COM INPUT 2 3 4 Hz Duty EMOS spol s r o Lipnick 2844 750 02 P erov...
Page 197: ...197...
Page 198: ...198...
Page 199: ...199...