2

GB | Digital Multimeter

Multimeter MD-220 is a compact, battery-powered, user-friendly 

device for measuring alternating and direct current, voltage and 

electrical resistance, for testing diodes and transistors and for sound 

testing of conductivity. The multimeter was designed in accordance 

with the IEC-61010 standard regarding electronic measuring devices 

in the category (CAT III 600 V), 2nd degree of pollution. 

The CAT III category is used to measure circuits powered by a fixed 

output power supply, such as a relay, socket, switchboards, power sup-

plies, short branching circuits and lighting systems in large buildings.

Safety Instructions

•  Read this manual thoroughly before using the multimeter.

•  Always check the connection of testing conductors and measuring 

range settings.

•  Do not exceed the maximum input limits:

•  for AC and DC voltage: 600 V

•  for alternating and direct current: 10 A

•  Before changing the measuring range (function), disconnect the 

conductors from the measured circuit.

Main Features

LCD display: maximum displed value 1999 (3 and 0.5 digits) with 

automatic polarity indication

Measuring method: dual – reacts to the pulse leading and trailing 

edge

Maximum standard mode: 500 V DC/AC ms

Reading frequency: approx. 2–3 readings per second

Operating temperature: 23 °C ±5 °C

Temperature range: operating 0 °C to +40 °C; storage –10 °C to 

+50 °C

Power supply: 9 V battery (1604 or 6F22). The status of the battery 

is indicated by the 

 icon on the left side of the display

Dimensions and weight: 85 × 165 × 32 mm, 250 g (including the 9 

V battery)

Accessories: Instruction manual, measuring tips

Operating Procedure

a.  If the device does not turn on (the display does not light up) or 

if the 

 symbol appears on the display upon turning on the 

device, the battery is completely drained. Replace the battery.

b.  When measuring quantities for which a 

 symbol is displayed 

below the measuring tip connection sockets, do not exceed the 

measuring range (risk of damaging the device).

c.  If you do not know the approximate voltage or current value 

beforehand, set the measuring range to maximum and grad-

ually decrease it as you measure.

d.  If the measuring range is exceeded (the display shows the 

number "1"), switch to a higher range.

e.  Avoid contact with high voltage.

Measuring Direct (DC) Voltage

1.  Connect the black measuring tip to the socket labelled as "COM" 

and the red conductor to the socket labelled as "V/Ω".

2.  Switch to the function marked as V

. Choose a measuring range 

and place the measuring tips where you wish to measure DC volt-

age. The voltage value and polarity will be shown on the display.

Follow points c), d), e) of the Operating Procedure!

Range

Accuracy

Deviation

200 mV

±0.5 % of reading 

± 1 counts

100 μV

2 V

±0.5 % of reading 

± 3 counts

1 mV

20 V

10 mV

200 V

100 mV

600 V

±0.8 % of reading 

± 2 counts

1 V

Input impedance: 10 MΩ for all ranges

Overload protection: 250 V rms of peak voltage at 200 mV and 

600 V rms in all other ranges.

Measuring Alternating (AC) Voltage

1.  Connect the black measuring tip to the socket labelled as "COM" 

and the red conductor to the socket labelled as "V/Ω".

2.  Switch to the function marked as V~. Choose a measuring range 

and place the measuring tips where you wish to measure AC 

voltage. The measured values will be displayed.

Follow points c), d), e) of the Operating Procedure!

Range

Accuracy

Deviation

2 V

±0.8 % of reading 

± 3 counts

1 mV

20 V

10 mV

200 V

100 mV

600 V

±1.2 % of reading 

± 3 counts

1 V

Input impedance: 10 MΩ for all ranges

Frequency range: 40 Hz to 400 Hz

Overload protection: 600 V of peak voltage in all ranges

Measuring Direct Current

1.  Connect the black measuring tip to the socket labelled as "COM" 

and the red conductor to the socket labelled as "mA". To measure 

maximum current (up to 10 A), connect the red conductor to the 

socket marked as 10 A.

2.  Switch to the function marked as A

. Choose a measuring range 

and place the measuring tips where you wish to measure DC cur-

rent. The current value and polarity will be shown on the display.

•  Maximum input current of 200 mA or 10 A depends on which 

socket the red measuring tip is plugged into. Excessive current 

will destroy the fuse. In that case, it has to be replaced. The 10 

A range is not protected by a fuse and can be measured for a 

maximum of 15 s, maximum voltage drop of 200 mV.

Follow points c), d) of the Operating Procedure!

Range

Accuracy

Deviation

2 mA

±0.8 % of reading 

± 1 count

1 μA

20 mA

10 μA

200 mA

±1.2 % of reading 

± 1 count

100 μA

10 A

±2 % of reading 

± 5 counts

10 μA

Overload protection: F 0.2 A/250 V fuse, (10 A range without fuse)

Maximum input current: 10 A, 15 seconds

Measuring Alternating Current

1.  Connect the black measuring tip to the socket labelled as "COM" 

and the red conductor to the socket labelled as "mA". To measure 

maximum current (up to 10 A), connect the red conductor to the 

socket marked as 10 A.

2.  Switch to the function marked as A~. Choose a measuring range 

and place the measuring tips where you wish to measure AC cur-

rent. The current value and polarity will be shown on the display.

•  Maximum input current of 200 mA or 10 A depends on which 

socket the red measuring tip is plugged into. Excessive current 

will destroy the fuse. In that case, it has to be replaced. The 10 

A range is not protected by a fuse and can be measured for a 

maximum of 15 s, maximum voltage drop 200 mV.

Follow points c), d) of the Operating Procedure!

Range

Accuracy

Deviation

2 mA

±1.2 % of reading 

± 3 counts

1 μA

20 mA

10 μA

200 mA

±2 % of reading 

± 3 counts

100 μA

10 A

±3 % of reading 

± 7 counts

10 mA

Overload protection: F 0.2 A/250 V fuse, (10 A range without fuse)

Maximum input current: 10 A, 15 seconds

Frequency: 40 Hz to 400 Hz

Maximum voltage drop: 200 mV

Indicates average value of sine waves.

Summary of Contents for M2092

Page 1: ...er CZ Digit ln multimetr SK Digit lny multimeter PL Multimetr cyfrowy HU Digit lis multim ter SI Digitalni multimeter RS HR BA ME Digitalni multimetar DE Digitales Multimeter UA RO Multimetru digital...

Page 2: ...V 20 V 10 mV 200 V 100 mV 600 V 0 8 of reading 2 counts 1 V Input impedance 10 M for all ranges Overload protection 250 V rms of peak voltage at 200 mV and 600 V rms in all other ranges Measuring Alte...

Page 3: ...es may reach groundwater and subsequently food chain where it could affect human health Emos spol s r o declares that the MD 220 is in compliance with the essential requirements and other relevant pro...

Page 4: ...e okruh je bez energie a e v echny kapacitn jednotky jsou vybity Rozsah P esnost Odchylka 200 0 8 z rozsahu 3 slice 0 1 2 k 0 8 z rozsahu 1 slice 1 20 k 10 200 k 100 2 M 1 000 20 M 1 z rozsahu 2 slic...

Page 5: ...vi body c d e Pracovn ho postupu Rozsah Presnos Odch lka 2 V 0 8 z rozsahu 3 slice 1 mV 20 V 10 mV 200 V 100 mV 600 V 1 2 z rozsahu 3 slice 1 V Vstupn impedancia 10 M pre v etky rozsahy Frekven n rozs...

Page 6: ...do pomiaru obwod w i wyposa e nia zasilanego z instalacji u o onej na sta e takiego jak przeka niki gniazdka panele rozdzielcze rozdzielnice zasilaj ce i kr tkie obwody oraz systemy o wietlenia w du...

Page 7: ...onego COM a przew d czerwony do zacisku oznaczonego V czer wony przew d pomiarowy pod czamy do 2 Prze cznikiem wybieramy funkcj oznaczon Dobieramy za kres pomiarowy i pod czamy przewody pomiarowe do p...

Page 8: ...r cs csot a V jelz s csatlakoz ba 2 Kapcsolja a k sz l ket a V jel funkci ra V lassza ki a m r si tartom nyt majd rintse a m r cs csokat oda ahol az egyen ram fesz lts get m rni k v nja A fesz lts g r...

Page 9: ...iztos t kot egy m sik azonos m ret F 0 2 A 250 V biztos t kra kell cser lni A k sz l ket korl tozott fizikai rz kszervi vagy ment lis k pess g szem lyek bele rtve a gyermekeket illetve a k sz l k hasz...

Page 10: ...na funkcijo ozna eno kot A Izberite merilni obseg in merilne konice priklju ite na mesto kjer elite meriti izmeni ni tok Polarnost se bo prikazala v istem asu kot tok Maksimalni vstopni tok 200 mA al...

Page 11: ...Ako se ure aj ne uklju uje zaslon se ne uklju uje ili ako se na zaslonu prilikom uklju ivanja poka e simbol baterija je posve ispra njena Zamijenite bateriju b Kod mjerenja koli ina za koje se prikaz...

Page 12: ...osigura a Za zamjenu osigura a uklonite cijeli stra nji poklopac i zamijenite osigura drugim osigura em od F 0 2 A 250 V iste veli ine Nije predvi eno da ovaj ure aj upotrebljavaju osobe uklju uju i...

Page 13: ...ist Durch berstrom wird die Sicherung zerst rt In diesem Fall muss diese ersetzt werden Der Bereich von 10 A ist nicht durch eine Sicherung gesch tzt und die Messung kann f r max 15 s erfolgen maximal...

Page 14: ...d legen Sie eine neue hinein Es wird empfohlen GP Batterien zu verwenden Wechsel der Sicherung Nehmen Sie beim Sicherungswechsel die komplette hintere Abde ckung ab und tauschen Sie die Sicherung F 0...

Page 15: ...5 10 A F 0 2 A 250 10 A 10 A 15 1 COM a A 10 10 2 A 200 10 10 15 200 c d 2 A 1 2 3 1 A 20 A 10 A 200 A 2 3 100 A 10 A 3 7 10 mA F 0 2 A 250 10 A 10 A 15 40 400 200 1 COM a V 2 1 1 200 0 8 3 0 1 2 k 0...

Page 16: ...e i conecta i sondele de m surare la locul n care ve i m sura tensiunea continu Se va afi a valoarea tensiunii i concomitent cu ea i polaritatea n continuare vezi punctele c d e ale Modului de lucru G...

Page 17: ...te privind utilizarea aparatului de c tre persoana responsabil de securitatea acestora Trebuie asigurat supravegherea copiilor pentru a se mpiedica joaca lor cu acest aparat Nu arunca i consumatorii e...

Page 18: ...a vesties srov 200 mA arba 10 A priklauso nuo to kur lizd ki tas raudonas matavimo antgalis D l pernelyg didel s srov s saugiklis gali perdegti Tokiu atveju reik s j pakeisti 10 A intervalas neapsaugo...

Page 19: ...in t aptuveno sprieguma vai str vas v rt bu iestatiet maksim lo m r anas diapazonu un m rot pak pe niski samaziniet to d Ja m r anas diapazons tiek p rsniegts displej tiek par d ts skaitlis 1 p rsl dz...

Page 20: ...omain tu dro in t ju no emiet visu aizmugur jo v ku un nomai niet dro in t ju ar citu t da pa a izm ra F 0 2 A 250 V dro in t ju o ier ci nav at auts lietot person m tostarp b rniem ar ierobe ot m fiz...

Page 21: ...nane m teotsik hendatakse Liialt k rge vool rikub kaitsme Sellisel juhul tuleb see asendada 10 A vahemikku kaitsmed ei kaitse ning seda saab m ta maksimaalselt 15 s maksimaalne pingelangus on 200 mV J...

Page 22: ...osk las di rektiivi n uete ja muude s tetega Seda seadet tohib ELi riikides vabalt kasutada Vastavusdeklaratsioon on osa kasutusjuhendist ja see on leitav ka kodulehel http www emos eu download BG MD...

Page 23: ...0 A 3 7 10 mA F 0 2 A 250 V 10 A 10 A 15 40 Hz 400 Hz 200 mV 1 COM V 2 1 1 200 0 8 3 0 1 2 k 0 8 1 1 20 k 10 200 k 100 2 M 1 000 20 M 1 2 10 k 250 V DC ms 1 COM V 2 1 COM V 2 50 hFE 1 PNP NPN 2 PNP NP...

Page 24: ...em obmo ju Republike Slovenije 9 Proizvajalec zagotavlja proti pla ilu popravilo vzdr evanje blaga nadomestne dele in priklopne aparate tri leta po poteku garancijskega roka 10 Naravna obraba aparata...

Reviews: