63
Handheld Probe Scan Probes Tab
Select and activate the scan area that you want to manually measure. Note
that you are not selecting probes on the EMxpert but rather an array of
measurements that you will do with the handheld probe. The EMxpert software
will organize the data into an equivalent array for display but this will only be
as accurate as the positioning of the probe. The number of active probes is the
number of manual probe measurements that you are required to complete. You
can activate the probes by either left click on individual cells or left click and
drag the mouse and choose Select Cells or click Select All to activate all of the
probes. Green indicates the activated probes.
You can deactivate the probes either left click on already selected cells or left click and drag the mouse
and choose Unselect Cells or click Select None to deactivate all of the probes.
Using the wheel of the mouse will adjust the transparency of the JPG file to help set the probes.
Overlay Editor
Please refer to
page 40
.
EMxpert Software
Handheld Probe Scan
Note:
Importing design
file or the picture of
the DUT is highly
recommended.
Please refer to
Overlay Editor
section.
Note:
The Handheld Probe Scan node is not using the patented scanner to
measure. You do not need to position the DUT on the scanner but you do
need the scanner connected to the adaptor.
Note:
The handheld scan node will organize the measurements into the correct physical
layout if the probes position is correct for each manual trigger. Please refer to
page 65
.