81
Error Messages
Possible Causes
Action
Failed to read license
Driver cannot read license on the adaptor
because of data transfer error.
1) Check the connection
between the adaptor and the
PC.
2) Try one more time.
3) If you get the same error
message again, the license
should be updated. Please
contact
EMSCAN
Technical
Support.
Failed to write license
Driver cannot write license to adaptor
because of data transfer error.
1) Check the connection
between the adaptor and the
PC.
2) Try one more time.
3) If you get the same error
message again, please contact
EMSCAN
Technical Support.
Error #338: The requested
bandwidth is outside the capabilities
of the spectrum analyzer. <detailed
bandwidth information>
The spectrum analyzer is not capable
of running with the requested resolution
bandwidth. The <detailed bandwidth
information> details what has been
requested and the bandwidth capabilities
of the attached spectrum analyzer.
Change the RBW (resolution
bandwidth) in the settings to a
supported value and rerun the
scan.
Error #340: The requested
frequency range is outside the
capabilities of the spectrum analyzer
<detailed range information>
The spectrum analyzer is not capable of
running with the requested frequencies.
The <detailed range information> details
what has been requested and the
frequency capabilities of the attached
spectrum analyzer.
Change the frequency limits
in the settings to supported
values and rerun the scan.
The start frequency is lower than the
licensed low frequency <detailed
range information>
The frequency setting is out of the
licensed frequency range
Change the lower frequency
limit in the settings to a
supported value and rerun the
scan. If you wish to change
the licensed frequency limits
please contact
EMSCAN
Technical Support.
The stop frequency is higher than the
licensed high frequency <detailed
range information>
The frequency setting is out of the
licensed frequency range
Change the upper frequency
limit in the settings to a
supported value and rerun the
scan. If you wish to change
the licensed frequency limits
please contact
EMSCAN
Technical Support.
There is a problem with the JPG file,
EMxpert can’t read it.
File is corrupted or it is not a JPG format
file.
There is a problem with the image
file, EMxpert can’t read it.
File is corrupted or it is not a Gerber
format file.
EMxpert Software