11
ERX+ Connection Overview
Item Number
Description
1
Scanner LED light shows three stages:
Red: Boot mode
Orange: Initiation mode (checking adaptor’s firmware)
Green: User mode (adaptor is connected and ready). The LED will be
blinking during the measurements.
2
ON / OFF switch. When ON the switch is lit
3
RF In port (input from external). You can connect external devices such as
a handheld probe or automated probe, and amplifier, attenuator, passband
filter etc. as long as the power into the RF In port does not exceed
-10 dBm
4
RF Out port (output directly from scanner). You can connect external
devices such as amplifier, attenuator, passband filter etc. to condition the
signal directly from the scanner before its input into the built-in spectrum
analyzer
5
Ethernet cable port
6
Universal power supply port
7
Fuse to provide protection against excess current
Specs: 3A / 250V, size: 0.177” Dia x 0.570” L (4.50 mm x 14.48 mm)
Recommended fuse: 0229003.HXP by Littelfuse Inc.
(Please refer to
Appendix: Fuse Replacement
)
8
Pinhole for the factory settings reset
About EMxpert
ERX+ Connection Overview
Caution:
The surface of the scanner is made of Gorilla
®
glass. It may shatter if a sharp object is
dropped from a distance.
7
8
6
1
2
3
4
5