System Setup (continued)
Caution:
Limit electrosurgical generator power settings as noted. Higher settings may
result in spurious insulation failure alarms and/or insulation breakdown.
ESU
Mode and Power Limitation (Watts)
Conmed System
System 5000
™
Use Lap-Specialty Mode Only: Max Power = 80
Covidien
Force Triad
™
All Modes: Max Power = 80
Covidien Force
FX
™
or FX-C
™
Med COAG: Max Power = 75
High COAG: Max Power = 60
All Other Modes: Max Power = 80
Covidien
FT10™
All Modes: Max Power = 80
Erbe
VIO 300 D
™
Dry Cut: Max Eff ect = 7, Max Power = 50
COAG Spray Eff ect 2: Max Power = 25
All Other Modes: Max Power = 50
Megadyne
MegaPower™
COAG 1: Max Power = 50
All Other Modes: Max Power = 80
Whenever the patient return electrode is connected to the EndoShield 2, the device
operates continuously.
System Setup (continued)
1. Connect the EndoShield 2 to the electrosurgical generator.
2. Connect the patient return electrode to the EndoShield 2.
illuminates RED until an AEM instrument is connected.
If no symbols illuminate, see the
Troubleshooting and System Maintenance
section.