background image

IM727 Rev A, Page 4 of 9 

 

 

®

unwanted static strain, resulting in possible zero shift during a shock measurement. The surface on 
which  the  accelerometer  is  mounted  should  have  a  surface  roughness  of  32  micro  inches  rms  or 
better and a surface flatness tolerance of 0.0003”.    

 

For use up to 60

o

C: Aremco Crystalbond

TM  

509 is recommended, following this mounting procedure: 

o

 

Apply a moderate amount of Crystalbond

TM  

509 on the bottom of the Model 727 accelerometer 

o

 

Use a toothpick and spread the wax to form a thin, uniform layer covering area of contact, and 
immediately position accelerometer on mating surface 

o

 

Press down and hold accelerometer firmly in place for 30 seconds while the wax sets.  Allow wax 
to set for at least 1 minute for maximum strength. 
 

   

Cyanoacrylate (super glue) is not recommended as a temporary adhesive unless an appropriate solvent 

   

(such as acetone) is used to weaken the glue joint before attempting to remove the accelerometer. The 

 

solvent should be swabbed around the base of the accelerometer; under no circumstances should the 

 

accelerometer  be soaked or immersed in any liquid. Similarly, if wax is used as a temporary adhesive, 

 

always remove the unit using the appropriate amount of heat. Trying to remove the accelerometer by 

 

shearing or twisting will excite the resonance and could cause permanent damage to the accelerometer. 

 

Mounting to low strength material (such as aluminum) is not recommended. Additionally, adhesive 
must always be used between the transducer and any low-strength mounting surface. 
 

CABLE HANDLING

 

 

The 727 series of drop-test accelerometers are supplied with integral cables, which are terminated 
in pigtails. The user can select an appropriate end-connector. 

 

There are two important considerations regarding the integral cable when installing the 
accelerometer. First, since the cable will be exposed to the same high level shocks as the 
accelerometer, care needs to be taken to prevent damage to the cable. Relative motion between the 
cable and the accelerometer can cause cable failure and/or a significantly shortened life. Second, the 
signal voltage on the leads of the cable will be at a very low level and any excessive movement of 
the cable could cause noise on the signal lines. Because of the above factors, it is important that the 
following cable recommendations be observed. 

 

If possible, the mounting preparation should allow for the cable to be routed perpendicular to the 
primary shock direction to reduce the amount of tensile stress on the cable. To strain relieve the 
cable close to the accelerometer, form a small bend (~¼” radius) in the cable within two inches of 
the case and then tack the cable to the mounting surface with tape. Next, completely secure the cable 
between the accelerometer case and the tape to the mounting surface by encapsulating with a 
silicone RTV. A recommended silicone RTV is Loctite® item number 37463, an air curing RTV 
which will be tack-free in 15-20 minutes, test-ready in roughly an hour and fully cured in 24 hours 
(assuming 70ºF and 50% relative humidity). In routing the remaining portion of the cable to the 
signal conditioner, it is important that there be sufficient slack in the cable, i.e. the cable should not 
be pulled tight between the test specimen and the signal conditioner. 
 

Summary of Contents for 727

Page 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Page 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Page 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Page 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Page 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Page 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Page 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Page 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Page 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Reviews: