background image

IM727 Rev A, Page 7 of 9 

 

 

®

 

 

OPERATING PRECAUTIONS

 

 

While  these  drop-test  accelerometers  can  be  extremely  rugged,  they  can  also  be  extremely  fragile 
under very specific operating conditions, namely under resonance excitation. Due to the extremely 
high  levels  of  shock  encountered  and  the  potential  for  resonance  excitation,  the  user  should 

read 

this section carefully 

before operating. 

 

RESONANT FREQUENCY EXCITATION 

 

The 727 series utilizes an undamped sensing element which means its resonant frequency is excited 
more easily as compared to a damped device.  Additionally, the 727 series does not incorporate any 
mechanical overtravel stops, which means the displacement of the proof mass is not restricted. Since 
the amplification factor for an undamped accelerometer can be up to 100, if there is sufficiently high 
frequency content in the shock event, the excessive amplitude of the seismic mass can be well over 
the overrange capability and cause permanent damage to the accelerometer.   

 

Another thing to note is that each seismic element used on the 727 series includes two masses, each 
with  its  own  resonant  frequency.  If  both  resonant  frequencies  are  excited,  a  beat  frequency  may 
appear in your data that will be at a frequency equivalent to the difference between the two resonant 
frequencies. Although the beat frequency will be apparent in the time history it will not translate to 
the Fourier transform. Although resonance excitation cannot always be avoided, it possible for the 
user  to  take  measures  to  reduce  the  amount  of  high-frequency  energy  the  accelerometer  will  be 
subjected to.  

 

BANDWIDTH AND AMPLITUDE LINEARITY 

 

Because  the  727  series  is  undamped  and  does  not  use  mechanical  over  travel  stops,  accurate 
measurements are permitted over a wide frequency and amplitude range. Undamped accelerometer 
data  are  typically  considered  accurate  (±5%)  to  one-fifth  of  the  resonant  frequency  and,  because 
there  are  no  overtravel  stops  to  limit  the  displacement  of  the  proof  mass,  reasonable  amplitude 
linearity can be observed well beyond the rated full-scale range. 

 

DATA ACQUISITION AND ANTI-ALIASING PRECAUTIONS 

 

With any shock testing it is always best practice to use recording equipment with enough bandwidth 
to resolve the highest frequency component present, which is usually the resonant frequency of the 
accelerometer.  Even  with  the  excitation  of  resonance,  if  linear  filtering  is  used  and  the  amplitude 
during resonance does not cause saturation to the signal conditioning (or go beyond the linear range 
of the damped accelerometer), the high amplitudes experienced during resonance will not degrade 
the accuracy of the actual shock data recorded at lower frequencies. 

 

The  user  should  also  be  aware  of  aliasing  and  its  effect  on  the  data.  A  simple,  and  by  no  means 
complete, explanation of aliasing is as follows. Aliasing occurs during digitization when you have 
energy  in  the  analog  time  history  at  a  frequency  higher  than  one-half  of  the  sampling  rate  (the 

Summary of Contents for 727

Page 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Page 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Page 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Page 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Page 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Page 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Page 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Page 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Page 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Reviews: